Key Techniques for Analyzing Samples Accurately By Using Atomic Absorption Spectrophotometer (AAS)

0
8كيلو بايت

In the realm of analyzing test substances, the preliminary treatment of samples presents a complex challenge. This aspect is pivotal within the scope of analysis and detection efforts, constituting a significant facet for identifying the origins of detection inaccuracies. Consequently, this article aims to synthesize four fundamental sample pretreatment techniques employed in detection endeavors of Atomic Absorption Spectrophotometer (AAS) . Additionally, it delves into the foundational detection techniques for six categories of routine samples. These methodologies offer practicality and serve as valuable references for users of Atomic Absorption Spectrophotometers (AAS).

Double Beam DW-AA320N Atomic Absorption Spectrophotometer (1 Lamp Stand)

The Four Core Techniques for General Sample Pretreatment in AAS

Wet Digestion Method

For samples approximately ranging from 0.1000 to 0.5000g, prevalent mixed acids are utilized, with varying ratios as follows:

HNO3 : HClO3 = 5 : 1

HNO3 : H2SO4 = 5 : 1

HNO3 : HCl = 5 : 1

Pure HNO3

Note: Volatile (such as acetone, ether, ethanol, etc.) and flammable, explosive substances must be absent during the digestion process. While the wet digestion method is widely employed, its operational specifics will not be reiterated here.

Dry Ashing Method

Suitable for sample sizes typically ranging from 2.000 to 5.000g, this method involves careful treatment to prevent volatilization. The process entails placing the sample in a porcelain crucible, moistening it with a few drops of water, adding a small quantity of concentrated nitric acid, heating it gently to carbonize, and subsequently ashing it in a muffle furnace at around 550°C for 2 to 4 hours. Post-cooling, the residue is dissolved with additional acids plus HNO3 (1:1), which varies depending on the sample. The solution is then filtered, adjusted to volume, and readied in 10mL, 25mL, and 50mL portions for subsequent use.

High-Pressure Vessel Method (Utilizing Polytetrafluoroethylene Lid-Made Tanks)

When dealing with samples below 0.3000g, this method involves introducing 6mL of mixed acid and 1mL of HF (H2O2) to the sample. The lidded autoclave is then sealed, and heating is performed at 160°C for 5 hours. After cooling, the sample is filtered, adjusted to volume, and set aside for future use.

Microwave Digestion Method

In this approach, commonly employed mixed acids encompass:

HNO3 : HClO3

HNO3 : H2SO4

Pure HNO3

It is essential to select the appropriate acid for specific digestion depending on the sample. Readers are encouraged to make suitable choices as per their requirements.

Conventional AAS Analysis and Detection Approaches for Diverse Samples

Analysis of Pb, Cd, As, Mo, Cr, etc. (Graphite Furnace AAS Method)

Pb: Extract 1.0mL of the sample, dilute it to 10mL with 1% HNO3. Linear range: 0~20ng/mL, drying temperature: 80~100°C, ashing temperature: 200°C, atomization temperature: 1500°C.

Cd: Extract 1.0mL of the sample, dilute it to 10mL with deionized water. Linear range: 0.1~0.4ng/mL, drying temperature: 80~100°C, ashing temperature: 200°C, atomization temperature: 1800°C.

As: Extract 1.0mL of the sample, add 100μL of Ni (2mg/mL), dilute to 10mL with 1% HNO3. Linear range: 0~4ng/mL, drying temperature: 80~100°C, ashing temperature: 200°C, atomization temperature: 2200°C.

Mo: Prepare a 1.0mL sample diluted with 1% HNO3 to 10mL. Using Pd as a modifier, the linear range is 0-20ng/mL. Drying temperature: 80-100°C, ashing temperature: 1200°C, atomization temperature: 2600°C.

Cr: Sample 1mL, dilute to 100mL with deionized water, and proceed with appropriate testing. Linear range: 0~40ng/mL, drying temperature: 80~100°C, ashing temperature: 1000°C, atomization temperature: 2600°C.

Analysis and Testing of Se, K, Na, Ca, Mg, etc. in Plant Samples

Plant samples are subjected to washing, drying, fixation at 150°C for 10 minutes, subsequent drying at 70°C, and final comminution before analysis.

Se in forage grass: Extract 1mL of the sample, add 100μL of Ni(NO3)2 improver (3mg/mL), dilute to 10mL with 1% HNO3 and 0.1% Triton (1:1). Linear range: 0~20ng/mL, drying temperature: 80~100°C, ashing temperature: 1000°C, atomization temperature: 2400°C.

K and Na in plants: Weigh 0.2000g of the sample, add 10mL of mixed acid, extract in a constant-temperature water bath at 90°C, filter to 50mL volume, and proceed with suitable sampling and testing.

Ca and Mg in plants: Incinerate the sample at 550°C, dissolve with hydrochloric acid (1+1), adjust to 50mL volume (with 5% CsCl2 modifier 2 mL), and proceed with appropriate sampling and testing.

Beryllium Analysis in Mushrooms and Tea (Transverse Heating AAS Method)

Prepare the sample by washing, drying, fixation at 105°C, subsequent drying at 70°C, cooling, and crushing through a 60 mesh sieve.

Weigh 1.0000g of sample, add 15mL of HNO3, 1mL of H2SO4, digest at a low temperature, dilute to 25mL, and proceed with suitable sa

mpling for testing. Linear range for Be: 0~8ng/mL. Test conditions: drying temperature: 130°C, ashing temperature: 1500°C, atomization temperature: 2300°C.

Germanium Testing in Beverages (Utilizing Transversely Heated Flat Graphite Tubes)

Test outcomes indicate a linear range of 0~200ng/mL, using Ni(NO3)2 improver (50μg/mL) in a 1% HNO3 medium. Instrument conditions: drying temperature: 130°C, ashing temperature: 800°C, atomization temperature: 2000°C.

Drawings of AAS treating samples

Calcium Detection in Steel Slag (Typically Employing Flame AAS Method)

For samples above 200 mesh, weigh 0.0500g into a 50mL PTFE crucible. Dissolve it with a water:nitric acid:hydrofluoric acid ratio of 5:8:6. Add sulfuric acid, heat, and adjust volume for analysis. To mitigate niobium interference, include

Triton-100 (20%), Vc (0.1 M/L), and 2% HNO3 as improvers.

Determination of Metal Elements in Blood Samples

Human blood encompasses diverse trace constituents, including heavy metal elements. Various detection methods can discern these components, with AAS being particularly effective. Specific digestion methods are elaborated in sections "2, 3), (3)." For instance:

Cu determination in serum (flame AAS method): Dilute 0.8mL of serum with 1% HNO3 to 10mL for testing.

As determination in serum (graphite furnace AAS method): Dilute with 1% HNO3 and utilize Ni(NO3)2 as a matrix modifier.

Cd determination in serum (graphite furnace AAS method): Process 2mL of serum with HNO3 and H2O2 for analysis.

Fe and Cu determination in albumin (flame AAS method): Utilize 1mL of the sample with mixed acid, adjust volume, and proceed with appropriate sampling.

Germanium determination in whole blood (graphite furnace AAS method): Dilute 1.0mL of venous blood, utilize a matrix modifier mixture, and perform analysis at defined temperatures.

This comprehensive overview highlights essential techniques for accurate analysis and detection of various samples by using atomic absorption spectrometry.

البحث
الأقسام
إقرأ المزيد
الألعاب
Unlock the Ultimate Team with EA FC 25 Coins: Buy FIFA Coins for Sale Now!
Unlock the Ultimate Team with EA FC 25 Coins: Buy FIFA Coins for Sale Now! The gaming landscape...
بواسطة Casey 2025-05-16 20:30:24 0 2كيلو بايت
Networking
Split Type Parking Cooler Market Industry Overview and Growth Opportunities, Forecast Period 2025-2034
The Split Type Parking Cooler Market has emerged as a key segment in the automotive and cooling...
بواسطة mrfrmarket 2025-01-14 05:02:51 0 3كيلو بايت
الألعاب
Crédit FC 25 : Maximisez Vos Avantages dans FIFA Ultimate Team 25
Crédit FC 25 : Maximisez Vos Avantages dans FIFA Ultimate Team 25 Dans l'univers...
بواسطة Casey 2025-07-21 16:52:31 0 984
الألعاب
**Titel: "EA FC 26 Coins Kaufen: Die besten Tipps für den Erwerb von FC26 Coins"**
EA FC 26 Coins Kaufen: Die besten Tipps für den Erwerb von FC26 Coins In der spannenden...
بواسطة Casey 2025-08-01 00:45:31 0 824
أخرى
Europe Off-Road Equipment Market Grows in Construction & Mining Sectors by 2034 Forecast
Europe Off-Road Equipment Market: Trends, Growth, and Future Outlook The Europe off-road...
بواسطة mrfrmarket 2025-06-17 04:47:37 0 1كيلو بايت